Amprobe Multitest-1000 Continuity-Tester Bedienungsanleitung Seite 14

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 122
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 13
CABLES USED FOR THE TEST If the cables are changed or extended be sure and
recalibrate the instrument. During a continuity
test, if the resistance value free of calibration (that
is the resistance value less the calibration offset
value) is negative, the symbols
ν as well as
blinking CAL are displayed (refer to 5
th
screen
paragraph 4.1.3).
WARNING
Never disconnect the test leads while the word “Measuring” is being
displayed.
4.1.1.1. Procedure to cancel calibration parameters and cal symbol
LOW
ν> 5
cal
Measuring
To cancel the calibration parameters (and
the symbol CAL) it is necessary to perform a
calibration procedure with a resistance
greater than 5 (for example with the test
leads disconnected). When a cancellation is
completed this screen is displayed first,
followed by the screen below:
Message >5
:
means that the
instrument
detected a
resistance greater
than 5 therefore
it will proceed
with Reset
procedure.
LOW
es
cal
The screen alongside is displayed for 2
seconds, after which the instrument emits a
long beep and then displays the default
screen relative to the LOW test under
AUTO mode without the symbol CAL.
Message rES:
means that the
instrument
cancelled
(RESET) the
calibration
parameter
Seitenansicht 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 121 122

Kommentare zu diesen Handbüchern

Keine Kommentare