Ohms
RANGE Accuracy
400.0
0.8% + 6d
4.000k, 40.00k, 400.0k
0.6% + 4d
4.000M
1.0% + 4d
40.00M
2.0% + 4d
Open Circuit Voltage : 0.4VDC typical
Audible Continuity Tester
Open Circuit Voltage: 0.4VDC typical
Range: 400.0; Accuracy: 1.5% + 6d
Audible threshold: between 10 and 120.
Diode Tester
Open Circuit Voltage Test Current (Typical)
< 1.6 VDC 0.4mA
Capacitance
RANGE
1)
Accuracy
2) 3)
500.0nF, 5.000µF, 50.00µF, 500.0µF,
3000µF
3.5% + 6d
1)
Additional 50.00nF range accuracy is not specified
2)
Accuracies with film capacitor or better
3)
Specified with battery voltage above 2.8V (approximately half full battery). Accuracy
decreases gradually to 12% at low battery warning voltage of approximately 2.4V
DCA Current (Clamp-on)
RANGE Accuracy
1) 2)
400.0A
0.0A ~ 400.0A 1.5% + 4d
1000A
400A ~ 800A 1.5% + 4d
800A ~ 900A 2.0% + 4d
900A ~ 1000A 5.0% + 30d
1)
Induced error from adjacent current-carrying conductor: < 0.01A/A
2)
Zero
mode must be applied to zero the unit before measurement to ensure
accuracy. This is particularly important below 20A.